Requalifying a sterilizer means running a fresh test series before it goes back to work: three consecutive spore tests (biological indicators) for each cycle type you use, plus three Bowie-Dick tests on a pre-vacuum steam sterilizer (CDC, 2008 guideline for disinfection and sterilization, sterilizing practices).

The series is due after installation, a relocation, a redesign, a major repair or a sterilization failure (CDC, 2008 guideline for disinfection and sterilization, sterilizing practices). Where CDC's own texts disagree — empty chamber or fully loaded — follow the instructions from the maker of your sterilizer and record which state each cycle ran in.

When a sterilizer must be requalified

The five triggers

“All steam, ETO, and other low-temperature sterilizers are tested with biological and chemical indicators upon installation, when the sterilizer is relocated, redesigned, after major repair and after a sterilization failure has occurred”

Source: CDC/HICPAC, guideline for disinfection and sterilization in healthcare facilities (2008), sterilizing practices.

In plain words: steam, ethylene oxide (ETO) and other low-temperature sterilizers all get biological and chemical indicator tests at five moments — installation, relocation, a redesign, a major repair, and after a sterilization failure (CDC, 2008 guideline for disinfection and sterilization, sterilizing practices).

Two limits travel with that sentence:

What the test series looks like

The three cycles

“Three consecutive empty steam cycles are run with a biological and chemical indicator in an appropriate test package or tray. Each type of steam cycle used for sterilization (e.g., vacuum-assisted, gravity) is tested separately. In a prevacuum steam sterilizer three consecutive empty cycles are also run with a Bowie-Dick test.”

Source: CDC/HICPAC, guideline for disinfection and sterilization in healthcare facilities (2008), sterilizing practices.

In plain words: three cycles in a row, each one carrying a biological and a chemical indicator in an appropriate test package or tray (CDC, 2008 guideline for disinfection and sterilization, sterilizing practices). Each type of steam cycle you use is tested on its own — e.g., vacuum-assisted and gravity get their own series (CDC, 2008 guideline for disinfection and sterilization, sterilizing practices). And on a pre-vacuum steam sterilizer, three empty cycles with a Bowie-Dick test are added to the series (CDC, 2008 guideline for disinfection and sterilization, sterilizing practices).

The Bowie-Dick cycles belong to a prevacuum steam sterilizer, so a gravity unit does not run them as part of requalification (CDC, 2008 guideline for disinfection and sterilization, sterilizing practices). If you are new to the test itself, the guide to which dental sterilizers need the Bowie-Dick test, and how often covers it.

Empty chamber or fully loaded: CDC's texts disagree

The three requalification cycles are not worded the same way in every CDC text. Here they are side by side, in CDC's own words.

Three requalification cycles: empty chamber or fully loaded
CDC textWhat it says
2003 dental guidelines “rechallenge the sterilizer with biological indicator tests in three consecutive empty chamber sterilization cycles after the cause of the sterilizer failure has been determined and corrected”Source: CDC, 2003 dental infection-control guidelines, recommendation VI.F.9.cChamber: empty
2024 dental page “If the repeat spore test result is positive, remove the sterilizer from service and do not use until it has been inspected or repaired and re-challenged with spore tests in three consecutive fully loaded chamber sterilization cycles.”Source: CDC, 2024 best practices for sterilization monitoring in dental settingsChamber: fully loaded

So the texts disagree: one runs the three cycles in an empty chamber (CDC, 2003 dental infection-control guidelines, recommendation VI.F.9.c), and the 2024 page runs them fully loaded (CDC, 2024 best practices for sterilization monitoring in dental settings). This page does not pick a side.

What to do about it: follow the instructions from the maker of your sterilizer, and record which chamber state each requalification cycle ran in. That is advice, not a quoted rule — this page's sources do not settle the disagreement. Dental Logbook records the chamber state on every requalification cycle and never chooses one for you.

When the sterilizer goes back in service

After a failed spore test, two paths lead back. The short one turns on the repeat test:

Repeat test negative

“If the repeat spore test is negative, and mechanical and chemical indicators are within normal limits, put the sterilizer back in service”

Source: CDC, 2003 dental infection-control guidelines, recommendation VI.F.8.c.

In plain words: a negative repeat spore test, with the mechanical and chemical indicators within normal limits, puts the sterilizer back in service (CDC, 2003 dental infection-control guidelines, recommendation VI.F.8.c).

Repeat test positive

“Do not use the sterilizer until it has been inspected or repaired or the exact reason for the positive test has been determined”

Source: CDC, 2003 dental infection-control guidelines, recommendation VI.F.9.a.

In plain words: after a positive repeat test, the sterilizer stays out of use until it has been inspected or repaired, or until the exact reason for the positive test has been determined (CDC, 2003 dental infection-control guidelines, recommendation VI.F.9.a). The three-cycle rechallenge then comes after the cause of the failure has been determined and corrected (CDC, 2003 dental infection-control guidelines, recommendation VI.F.9.c).

Back in service after requalification means the whole series is recorded as passed: three consecutive negative spore tests per cycle type, and on a pre-vacuum unit the three Bowie-Dick tests. A good log marks the sterilizer back in service only then, with the date and the name of the person who released it.

The whole failure path — the repeat test, the loads to recall, and the questions to ask — is in what to do when a dental sterilizer fails a spore test.

What to record

A good requalification record holds these. They are product defaults, not quoted rules.

Dental Logbook keeps these fields on one requalification record per sterilizer, with the chamber state on every cycle, and it never asks for patient names.

Where these sources are silent

This page is not legal advice; check your state board's current rules.