Weekly spore test calendar for 2027

Print one sheet per sterilizer and keep it by the sterilizer all year. Each week of 2027 has its own box for the spore test, the control from the same lot and implant-load tests — and the Florida sheet adds use time and the 30-day limit.

What the rules say

The four rules this calendar is built on

Each rule below is quoted from the source that states it, with the row ID this site keeps it under. Nothing on this page adds a rule the sources do not hold.

  • Test each sterilizer at least weekly with a biological indicator and a matching control, meaning the biological indicator and the control come from the same lot number (CDC, 2003 dental infection-control guidelines).

    “Monitor sterilizers at least weekly by using a biological indicator with a matching control (i.e., biological indicator and control from same lot number)”

    CDC, 2003 dental infection-control guidelines, Recommendation VI.F.6 · row C03-1

    “A spore test should be used at least weekly to monitor sterilizers.”

    CDC, 2016 summary of infection prevention practices · row SUM-1

    “A biologic indicator (i.e., spore test) is used at least weekly and with every load containing implantable items”

    CDC infection prevention checklist for dental settings · row CHK-1

  • Put a biological indicator in every sterilizer load that contains an implantable device, and verify the results before the implantable device is used, whenever possible (CDC, 2003 dental infection-control guidelines).

    “Use a biological indicator for every sterilizer load that contains an implantable device. Verify results before using the implantable device, whenever possible”

    CDC, 2003 dental infection-control guidelines, Recommendation VI.F.7 · row C03-3

  • In Florida, steam under pressure (e.g. autoclave) and chemical vapor sterilizers must have their efficacy verified by appropriate biological monitoring at least once every 40 hours (2,400 minutes) of use or at least once every thirty days, whichever comes first (Florida rule 64B5-25.003).

    “Adequacy of steam under pressure (e.g. autoclave) or chemical vapor sterilization must have their efficacy verified by appropriate biological monitoring at least once every 40 hours (2400 minutes) of use or at least once every thirty days, whichever comes first.”

    Florida rule 64B5-25.003(9) · row FL-1

  • In Florida, dry heat and ethylene oxide sterilizers must have their efficacy verified with appropriate biological monitoring every 120 hours of operation at sterilization parameters or every thirty days, whichever comes first (Florida rule 64B5-25.003).

    Use time is determined by multiplying the number of cycles by the individual cycle time (Florida rule 64B5-25.003).

    “Dry heat and ethylene oxide sterilizers must have their efficacy verified with appropriate biological monitoring every 120 hours of operation at sterilization parameters or every thirty days, whichever comes first. (Use time is determined by multiplying the number of cycles by the individual cycle time.)”

    Florida rule 64B5-25.003(9) · row FL-2

Side by side

Where the sources differ — and where they are silent

CDC's recommendation and Florida's rule set different schedules. This page shows both and marks what each one does not say.

The sources on this page do not agree on one schedule: CDC's recommendation is at least weekly (CDC, 2003 dental infection-control guidelines; CDC, 2016 summary of infection prevention practices), and Florida's rule counts hours of use and days instead (Florida rule 64B5-25.003).

This calendar, and Dental Logbook, mark a test due at the earliest of the limits, so one sheet answers both schedules. That choice of “earliest” is ours; no source quoted here asks for a calendar at all.

CDC's recommendation and Florida's rule, side by side
On these sheetsCDC's recommendationFlorida's rule
One box per week A spore test at least weekly for each sterilizer (CDC, 2003 dental infection-control guidelines). At least every 40 hours (2,400 minutes) of use or at least every 30 days, whichever comes first, for steam and chemical vapor; every 120 hours of operation or every 30 days, whichever comes first, for dry heat and ethylene oxide (Florida rule 64B5-25.003).
The implant-load box A biological indicator for every load that contains an implantable device, with results verified before the device is used, whenever possible (CDC, 2003 dental infection-control guidelines). Silent. The Florida rows quoted on this page set intervals and use time; they do not mention implant loads.
The control box A biological indicator with a matching control, meaning the biological indicator and the control from the same lot number (CDC, 2003 dental infection-control guidelines). Silent. The Florida rows quoted on this page do not mention a control.
Use time and 30 days Silent. The CDC rows quoted on this page do not count use time. Use time is determined by multiplying the number of cycles by the individual cycle time, and the test interval is 40 hours (2,400 minutes) of use or 30 days for steam and chemical vapor, 120 hours of operation or 30 days for dry heat and ethylene oxide, whichever comes first (Florida rule 64B5-25.003).

What the quoted sources do not say

  • They do not say which weekday a test must fall on. The boxes on these sheets are numbered weeks, Monday to Sunday; the office chooses the day.
  • They do not say what a calendar should look like. The layout, the columns and the box sizes are this page's design.
  • The Florida rows quoted here do not mention implant loads. The implant box on both sheets follows CDC's recommendation (CDC, 2003 dental infection-control guidelines).
  • They do not say how long to keep the filled sheets. This page sets no retention period.
Template

The 2027 calendar

One sheet per sterilizer. Fill it in on screen and print it, or print it blank and write it in by hand. Nothing you type on this page is stored or sent anywhere.

Two US Letter pages, landscape: weeks 1–25, then weeks 26–53.

Weekly spore test calendar 2027

One sheet per sterilizer

Weeks run Monday to Sunday. The first box is the week of Dec 28, 2026 and the last is the week of Dec 27, 2027, so every day of 2027 has a box.

Weekly spore test calendar 2027, weeks 1 to 25: one row per week, one sheet per sterilizer.
Week Date tested BI product and lot Control, same lot (result) Result Read by Implant load tests this week (load no., date, result)
1 · Dec 28 – Jan 3
2 · Jan 4 – 10
3 · Jan 11 – 17
4 · Jan 18 – 24
5 · Jan 25 – 31
6 · Feb 1 – 7
7 · Feb 8 – 14
8 · Feb 15 – 21
9 · Feb 22 – 28
10 · Mar 1 – 7
11 · Mar 8 – 14
12 · Mar 15 – 21
13 · Mar 22 – 28
14 · Mar 29 – Apr 4
15 · Apr 5 – 11
16 · Apr 12 – 18
17 · Apr 19 – 25
18 · Apr 26 – May 2
19 · May 3 – 9
20 · May 10 – 16
21 · May 17 – 23
22 · May 24 – 30
23 · May 31 – Jun 6
24 · Jun 7 – 13
25 · Jun 14 – 20

Scroll the table sideways to reach every column.

Weekly spore test calendar 2027, weeks 26 to 53: one row per week, one sheet per sterilizer.
Week Date tested BI product and lot Control, same lot (result) Result Read by Implant load tests this week (load no., date, result)
26 · Jun 21 – 27
27 · Jun 28 – Jul 4
28 · Jul 5 – 11
29 · Jul 12 – 18
30 · Jul 19 – 25
31 · Jul 26 – Aug 1
32 · Aug 2 – 8
33 · Aug 9 – 15
34 · Aug 16 – 22
35 · Aug 23 – 29
36 · Aug 30 – Sep 5
37 · Sep 6 – 12
38 · Sep 13 – 19
39 · Sep 20 – 26
40 · Sep 27 – Oct 3
41 · Oct 4 – 10
42 · Oct 11 – 17
43 · Oct 18 – 24
44 · Oct 25 – 31
45 · Nov 1 – 7
46 · Nov 8 – 14
47 · Nov 15 – 21
48 · Nov 22 – 28
49 · Nov 29 – Dec 5
50 · Dec 6 – 12
51 · Dec 13 – 19
52 · Dec 20 – 26
53 · Dec 27 – Jan 2

This page is not legal advice; check your state board's current rules.

For Florida offices

The Florida sheet: use time and 30 days

The same weekly boxes, plus the count Florida's rule asks for. The formula and a worked example are printed on the sheet.

When the next test is due

The date is the earliest of these three limits.

  1. 7 days after the last test — at least weekly (CDC, 2003 dental infection-control guidelines), counted here as 7 days.
  2. 2,400 use-minutes since the last test for steam and chemical vapor, or 7,200 use-minutes for dry heat and ethylene oxide — every 40 hours (2,400 minutes) of use or every 30 days, whichever comes first, and every 120 hours (7,200 minutes) of operation or every 30 days, whichever comes first (Florida rule 64B5-25.003).
  3. 30 days after the last test (Florida rule 64B5-25.003).

Use-minutes are counted as the logged duration of every cycle since the last test: use time is determined by multiplying the number of cycles by the individual cycle time (Florida rule 64B5-25.003).

Taking the earliest of the limits is how this calendar and Dental Logbook handle the two schedules. The sources set them for different offices.

Worked example

A steam sterilizer runs 8 cycles of 60 minutes a day. Its use time per day is 8 × 60 = 480 use-minutes, and 480 × 5 = 2,400 use-minutes.

8 cycles × 60 min = 480 use-min a day
2,400 ÷ 480 = 5 days
→ the 2,400-use-minute limit falls on day 5, before the 7-day and 30-day limits, so the test is due on day 5.

The formula is the one quoted above (Florida rule 64B5-25.003); the numbers are an example, not a rule.

Two US Letter pages, landscape: weeks 1–25, then weeks 26–53.

Weekly spore test calendar 2027 — Florida sheet

One sheet per sterilizer

Due at the earliest of 7 days (CDC, 2003 dental infection-control guidelines), 2,400 use-min since the last test for steam and chemical vapor or 7,200 for dry heat and ethylene oxide, or 30 days (Florida rule 64B5-25.003). Use-min = cycles × cycle time (Florida rule 64B5-25.003).

Florida sheet of the 2027 spore test calendar, weeks 1 to 25: weekly boxes with use time and days since the last test.
Week Date tested BI product and lot Control, same lot (result) Result Implant load tests Cycles Cycle time (min) Use-min this week Use-min since last test Days since last test
1 · Dec 28 – Jan 3
2 · Jan 4 – 10
3 · Jan 11 – 17
4 · Jan 18 – 24
5 · Jan 25 – 31
6 · Feb 1 – 7
7 · Feb 8 – 14
8 · Feb 15 – 21
9 · Feb 22 – 28
10 · Mar 1 – 7
11 · Mar 8 – 14
12 · Mar 15 – 21
13 · Mar 22 – 28
14 · Mar 29 – Apr 4
15 · Apr 5 – 11
16 · Apr 12 – 18
17 · Apr 19 – 25
18 · Apr 26 – May 2
19 · May 3 – 9
20 · May 10 – 16
21 · May 17 – 23
22 · May 24 – 30
23 · May 31 – Jun 6
24 · Jun 7 – 13
25 · Jun 14 – 20

Scroll the table sideways to reach every column.

Florida sheet of the 2027 spore test calendar, weeks 26 to 53: weekly boxes with use time and days since the last test.
Week Date tested BI product and lot Control, same lot (result) Result Implant load tests Cycles Cycle time (min) Use-min this week Use-min since last test Days since last test
26 · Jun 21 – 27
27 · Jun 28 – Jul 4
28 · Jul 5 – 11
29 · Jul 12 – 18
30 · Jul 19 – 25
31 · Jul 26 – Aug 1
32 · Aug 2 – 8
33 · Aug 9 – 15
34 · Aug 16 – 22
35 · Aug 23 – 29
36 · Aug 30 – Sep 5
37 · Sep 6 – 12
38 · Sep 13 – 19
39 · Sep 20 – 26
40 · Sep 27 – Oct 3
41 · Oct 4 – 10
42 · Oct 11 – 17
43 · Oct 18 – 24
44 · Oct 25 – 31
45 · Nov 1 – 7
46 · Nov 8 – 14
47 · Nov 15 – 21
48 · Nov 22 – 28
49 · Nov 29 – Dec 5
50 · Dec 6 – 12
51 · Dec 13 – 19
52 · Dec 20 – 26
53 · Dec 27 – Jan 2

This page is not legal advice; check your state board's current rules.

Using the sheets

How to use it

Four steps, from blank sheets to a year of weekly tests on paper.

  1. One sheet per sterilizer

    Write the office name, the sterilizer's label and the year at the top. Print a second set for each extra sterilizer.

  2. Fill the week's box after each test

    Date tested, the BI product and its lot, the control from the same lot with its result, the result, and your initials.

  3. Record implant loads in the same week

    Every load that contains an implantable device gets its own biological indicator (CDC, 2003 dental infection-control guidelines). Write those tests in the implant box of that week's row.

  4. In Florida, keep the count too

    Add the week's cycles and cycle time, then the use-minutes this week, the use-minutes since the last test, and the days since the last test. A test is due at the earliest limit.